X-Ray Preventive Maintenance Checklist

By MedBazaar Technical Team · Troubleshooting Guide

X-Ray System Preventive Maintenance Checklist

### روزانہ / Daily (Radiographer)

  • System startup normal
  • Tube warm-up protocol complete
  • Collimator light working
  • Bucky movement smooth
  • Image quality acceptable (test exposure agar protocol hai)
  • AEC working properly
  • PACS connectivity verified
  • No error messages
  • ### ہفتہ وار / Weekly (Radiographer + Biomedical)

  • Cable inspection (HV cable, detector cables)
  • Collimator alignment check
  • Table movement aur locks
  • Cassette/plate condition (CR systems)
  • Error log review
  • Disk space check
  • ### ماہانہ / Monthly (Biomedical Engineer)

  • kV accuracy check (kVp meter se)
  • mAs accuracy check
  • Timer accuracy
  • Collimation accuracy
  • AEC calibration check
  • Mechanical inspection: all moving parts
  • Electrical connections tighten karein
  • ### سالانہ / Annual (Certified Engineer + Physicist)

  • Full electrical safety test
  • Radiation output measurement
  • HVL (Half Value Layer) measurement
  • Dose calibration
  • Image quality comprehensive test
  • PNRA compliance verification
  • Regulatory certificate renewal
  • Tube performance assessment
  • All safety interlocks test
  • PNRA Requirements Pakistan

    Annual radiation survey mandatory hai. Licensed physicist ka certificate required hai for regulatory compliance. Non-compliance = facility closure risk.