Portable X-Ray Exposure Failure

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Portable X-ray exposure button press karne par X-ray fire nahi karta ya incomplete exposure hoti hai.

ممکنہ وجوہات / Possible Causes

  • Battery level insufficient for exposure
  • Exposure settings out of range
  • Tube fault
  • Generator fault
  • Ready state not achieved
  • تشخیص / Diagnosis Steps

  • Ready indicator light on hai check karein — unit ready state mein hai
  • Battery level check karein — exposure ke liye sufficient charge chahiye
  • kV aur mAs settings range mein hain verify karein
  • Error code note karein
  • حل / Fix

    Battery Level:

  • Min 80% battery recommended for reliable exposures
  • Fully charge karein
  • During exposure: avoid moving unit — current draw high hoti hai
  • Not Ready:

  • Ready indicator ka wait karein — capacitor charge hone do
  • Beep sound aaye tab exposure karein (agar applicable)
  • Settings:

  • Correct anatomical program select karein
  • Manual settings range mein hain verify karein
  • Engineer Required For

  • Tube fault diagnosis
  • Generator circuit repair
  • Exposure timing circuits
  • Estimated Fix Time: 15 min (operational) to engineer assessment (hardware)

    Difficulty: Intermediate

    🔧