مسئلہ / Problem

Portable X-ray battery jaldi discharge ho jati hai ya charge nahi hoti.

Battery Types

  • Sealed Lead Acid (SLA) — older units
  • Li-Ion — modern units (better performance)
  • NiMH — some older units
  • مسائل اور حل

    Rapid Discharge:

  • Battery age check karein — SLA: 2-3 years, Li-Ion: 3-5 years
  • Usage pattern: zyada exposures per charge = faster drain (normal)
  • Battery capacity test karein (engineer tools se)
  • Not Charging:

  • Charger output voltage check karein
  • Battery temperature check karein — hot battery charge nahi hoti
  • Battery completely dead: trickle charge 2-4 hours try karein
  • Memory Effect (NiMH batteries):

  • Full discharge phir full charge (conditioning)
  • 3-5 cycles repeat karein
  • Battery Replacement Guide

  • Model number note karein
  • Manufacturer ya compatible replacement lein
  • SLA replacement: relatively easy (user possible)
  • Li-Ion replacement: engineer recommended
  • New battery: 3-5 full charge cycles conditioning karein
  • Pakistan Tip

    Lead acid batteries Pakistan mein locally available hain — cost effective. Li-Ion ke liye authorized dealer se lein — cheap Chinese alternatives fire risk hain.

    Estimated Fix Time: Battery replacement: 30-60 min

    Difficulty: Basic to Intermediate

    🔧

    Cant Fix It Yourself?