Philips MP30 Calibration Guide

By MedBazaar Technical Team · Troubleshooting Guide

Calibration Requirements

### Mandatory Calibrations

IBP Zero Calibration - Every use se pehle

CO2 Zero - Daily ya module warm-up ke baad

Temperature Check - Annually

### IBP Zero Procedure

  • Patient connection close karein (stopcock atmospheric par)
  • MP30 mein: Measurements → IBP → Zero IBP
  • "Zeroed" confirmation wait karein
  • Stopcock patient par open karein
  • ### CO2 Calibration

  • Sensor room air mein expose karein (CO2 free)
  • Measurements → CO2 → Zero CO2
  • Span cal: certified gas (5% CO2) available ho to use karein
  • ### NIBP Verification

    Annual NIBP calibration engineer se karwain:

  • Reference manometer se comparison
  • Accuracy ±3 mmHg honi chahiye
  • All pressure ranges test karein
  • ### Service Calibration (Engineer Only)

  • ECG amplitude calibration
  • Respiration calibration
  • Full system performance test
  • Document results
  • Documentation

    Calibration records maintain karein:

  • Date
  • Technician name
  • Results
  • Pass/Fail
  • Next due date
  • Difficulty: Intermediate

    🔧