Mindray iPM 10 SpO2 Sensor Error

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

iPM 10 SpO2 reading nahi aa rahi ya "SpO2 Sensor Off" alarm aa raha hai.

ممکنہ وجوہات / Possible Causes

  • Sensor probe properly placed nahi
  • Patient ke nails par nail polish hai
  • Sensor cable damaged
  • Poor perfusion (patient specific)
  • تشخیص / Diagnosis Steps

  • Sensor position check karein — fingertip properly mein placed hai
  • Nail polish ya acrylic nails check karein
  • Dusri finger par try karein
  • Dusra SpO2 sensor try karein (agar available)
  • حل / Fix

  • Sensor reposition karein — light emitter aur detector aligned hone chahiye
  • Nail polish remove karein ya sensor sideways apply karein
  • Sensor cable inspect karein — kink ya damage
  • Ear probe ya forehead probe available ho to use karein
  • Sensor clean karein — blood ya contamination check
  • Engineer کب بلائیں

    SpO2 module replacement ya calibration ke liye engineer zaruri hai. Agar multiple sensors fail ho rahe hain to module fault hai.

    Estimated Fix Time: 5–15 minutes

    Difficulty: Basic

    🔧