Mindray iPM 12 SpO2 Error Guide

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

iPM 12 par SpO2 values inaccurate hain ya sensor errors frequent aa rahe hain.

ممکنہ وجوہات / Possible Causes

  • Sensor type mismatch (Nellcor vs Masimo vs Mindray)
  • Bright ambient light interfering
  • Poor peripheral circulation
  • Motion artifact
  • تشخیص / Diagnosis Steps

  • Sensor type check karein — iPM 12 specific sensor required
  • Ambient light check — direct sunlight ya bright OR lights
  • Patient circulation assessment
  • Sensor movement during measurement
  • حل / Fix

  • Compatible sensor use karein (Mindray SpO2 module ke liye Mindray ya compatible sensor)
  • Sensor cover se ambient light block karein
  • Warm extremity — circulation improve hogi
  • Motion: patient ko still rehne ko kahein ya different site try karein (earlobe, forehead)
  • Sensor Types Guide

  • Standard: Fingertip probe
  • Alternative: Ear clip probe (motion tolerance better)
  • Neonatal: Wrap sensor (premature babies ke liye)
  • Estimated Fix Time: 5–15 minutes

    Difficulty: Basic

    🔧