Mindray iPM 12 ECG Cable Fault

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

ECG waveform noisy hai, intermittent hai ya completely absent hai.

ممکنہ وجوہات / Possible Causes

  • Cable near connector broken internally
  • Lead wire snap connectors worn
  • Trunk cable damaged
  • EMI interference
  • تشخیص / Diagnosis Steps

  • Cable wiggle test: cable move karte waqt waveform change hoti hai to cable fault
  • Individual lead wires inspect karein
  • Trunk cable connections check karein
  • Nearby electrical equipment check karein — EMI source
  • حل / Fix

  • Faulty lead wire replace karein (individual wires replaceable hain)
  • Trunk cable replace karein agar damage visible hai
  • EMI source door rakhein (suction machines, electrocautery)
  • Proper grounding check karein
  • Cable Lifespan

    ECG cables typically 1-2 years with regular use. High-quality cables (Mindray OEM) longer last karte hain.

    Estimated Fix Time: 15–30 minutes

    Difficulty: Basic

    🔧