By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Battery backup kam ho gayi hai, charge nahi ho rahi ya battery alarms aa rahe hain.

ممکنہ وجوہات / Possible Causes

  • Battery age (>2 years)
  • Deep discharge damage
  • Charging circuit fault
  • Extreme temperature exposure
  • Battery Health Assessment

    Menu → Maintenance → Battery → Battery Info

  • Capacity: 80% se upar healthy
  • Cycle count: 300+ cycles = replacement consider karein
  • Voltage: normal range check karein
  • حل / Fix

  • Battery conditioning: full discharge phir full charge (2-3 cycles)
  • Battery replacement agar capacity <80%
  • Storage: 40-80% charge par store karein
  • Extreme heat/cold se bachein
  • Battery Replacement

    T5 battery typically 11.1V Li-Ion. Mindray authorized dealer se lein. Self-replacement possible hai lekin engineer se karwana recommended.

    Estimated Fix Time: 30 minutes

    Difficulty: Intermediate

    🔧