DR System Calibration Overview

DR detectors regular calibration require karte hain optimal image quality ke liye.

Calibration Types

Dark Field Calibration (Offset Cal)

  • Purpose: Electronic noise baseline remove karna
  • Frequency: Daily ya per session
  • Procedure: No exposure — detector covers hai
  • User level: Software mein "Dark Cal" ya "Offset Cal" button
  • Flood Field Calibration (Gain Cal)

  • Purpose: Detector response uniformity
  • Frequency: Weekly ya monthly
  • Procedure: Uniform X-ray exposure to detector (no patient, no grid)
  • User/engineer level
  • Defect Map Update

  • Purpose: Dead pixels/elements map karna
  • Frequency: Quarterly ya as needed
  • Engineer level
  • Daily Calibration Procedure

  • Software mein calibration mode enter karein
  • Detector properly position karein (no patient, clear field)
  • "Dark Cal" run karein — no exposure needed
  • Confirmation message wait karein
  • Normal use resume karein
  • When to Call Engineer

  • Calibration fail hoti hai
  • Artifacts persist despite calibration
  • Image quality baseline se significantly different
  • After any detector drop/impact
  • Image Quality Verification

    After calibration: flat field image lein — uniform grey, no bands ya patterns visible honi chahiye.