CR Reader Common Faults
By MedBazaar Technical Team · Troubleshooting Guide
CR Reader (Computed Radiography) Common Issues
1. Cassette Not Accepted / Jammed
*Cause:* Cassette damaged, dirty transport rollers
*Fix:* Cassette inspect karein — bent corners. Rollers clean karein. Jammed cassette: manual release procedure (model manual dekhen). Force na karein.
2. No Image After Reading
*Cause:* Plate not exposed, plate fogged, laser alignment
*Fix:* Plate exposure verify karein. Fog check: plate dark room mein expose karo — immediate reading. Laser: engineer required.
3. Image Artifacts (Lines/Bands)
*Cause:* Transport roller marks, dust on laser path, plate damage
*Fix:* Internal cleaning (engineer). Plate replace karein agar scratched.
4. Slow Reading Speed
*Cause:* Software issue, disk full, network bottleneck
*Fix:* Disk space check. Network speed check. Software restart.
5. Plate Erasure Incomplete
*Cause:* Erase lamp failing, erase time too short
*Fix:* Erase cycle complete hone do. Lamp replacement: engineer.
6. Software Not Communicating
*Cause:* Network issue, database error
*Fix:* Network check. Software restart. IT/engineer.
Maintenance Tips