Mindray DP-30 Probe Connector Fault

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Probe connector loose feel hota hai ya probe baar baar disconnect hota hai during exam.

ممکنہ وجوہات / Possible Causes

  • Connector locking mechanism worn
  • Socket pins loose
  • Connector housing cracked
  • Probe cable strain relief damaged
  • تشخیص / Diagnosis Steps

  • Connector lock mechanism inspect karein
  • Probe manually hold kar ke exam karein — stabilize hoti hai to connector issue
  • Socket mein loose movement check karein
  • حل / Fix

  • Connector locking lever replacement
  • Probe cable strain relief tape se temporarily secure karein
  • Socket assembly replacement (engineer required)
  • Engineer کب بلائیں

    Socket assembly replacement ke liye machine disassembly zaroori hai — engineer bulain.

    Estimated Fix Time: 1–2 hours

    Difficulty: Intermediate

    🔧