مسئلہ / Problem

Mindray DP-10 probe connect karne ke baad screen par "No Probe" ya "Probe Not Connected" error aata hai.

ممکنہ وجوہات / Possible Causes

  • Probe connector pins bent ya damaged hain
  • Probe socket mein dust ya moisture hai
  • Probe cable internally broken hai
  • Software ne probe recognize nahi kiya
  • تشخیص / Diagnosis Steps

  • Probe connector ko carefully inspect karein — bent pins check karein
  • Dusra probe available ho to test karein — issue probe mein hai ya machine mein
  • Connector area ko dry compressed air se clean karein
  • Machine restart karein aur dobara connect karein
  • حل / Fix

  • Bent pins ko needle-nose pliers se gently straighten karein
  • Connector cleaning ke liye isopropyl alcohol (70%) use karein
  • Probe cable kinks check karein — especially connector ke paas
  • Software update available ho to install karein
  • Engineer کب بلائیں

    Agar cleaning aur restart ke baad bhi probe detect nahi ho raha, probe internally damaged ho sakta hai. Certified engineer se inspection karwain.

    Estimated Fix Time: 30–60 minutes

    Difficulty: Intermediate

    🔧

    Cant Fix It Yourself?

    Book a certified biomedical engineer in your city. Fast response, verified professionals.

    Find an Engineer →