Skip to content
MarketplaceEngineersSellersCommunityBlogPricing
🔬UltrasoundMindray DP-20

Mindray DP-20 Probe Recognition Error

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

DP-20 connected probe ko recognize nahi karta aur "Probe Error" ya "Unknown Probe" message aata hai.

ممکنہ وجوہات / Possible Causes

  • Probe EEPROM chip faulty
  • Connector pins corrosion
  • Probe model machine software se incompatible
  • Probe connector damaged
  • تشخیص / Diagnosis Steps

  • Probe connector carefully inspect karein
  • Known working probe se test karein
  • Software version check karein — probe compatibility list dekhen
  • Connector cleaning karein
  • حل / Fix

  • Connector gold pins ko isopropyl alcohol se clean karein
  • Machine power cycle karein
  • Software update se naye probe models add ho sakte hain
  • Probe EEPROM replacement (factory service required)
  • Engineer کب بلائیں

    EEPROM replacement ya probe compatibility issues ke liye Mindray authorized service center se contact karein.

    Estimated Fix Time: 30–60 minutes

    Difficulty: Intermediate

    🔧

    Cant Fix It Yourself?

    Book a certified biomedical engineer in your city. Fast response, verified professionals.

    Find an Engineer →