GE Logiq P9 Probe Connector Error

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Logiq P9 probe connector error ya "Active Probe" selection issue.

ممکنہ وجوہات / Possible Causes

  • Multi-port confusion (P9 mein multiple probe ports hain)
  • Connector latch not fully engaged
  • Port-specific hardware fault
  • Probe PROM data corrupted
  • تشخیص / Diagnosis Steps

  • Correct port use kar rahe hain check karein (P9 mein ports labeled hain)
  • Probe connector click sound ke saath lock kiya hai verify karein
  • Dusre port par same probe try karein
  • Dusra probe same port par try karein
  • حل / Fix

  • Active probe software mein manually select karein
  • Connector latch clean aur lubricate karein (dry lubricant only)
  • Port cycling — probe unplug, 10 sec wait, reconnect
  • Port repair (engineer required)
  • Estimated Fix Time: 30–60 minutes

    Difficulty: Intermediate

    🔧