GE Logiq P9 C1-6 Probe Troubleshooting

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

C1-6 curved array probe specific issues — poor abdominal imaging ya probe errors.

ممکنہ وجوہات / Possible Causes

  • Probe face delamination
  • Cable strain near connector
  • Element failure (dead elements)
  • Frequency selection incorrect
  • تشخیص / Diagnosis Steps

  • Probe face visually inspect karein — bubbles ya separation visible hai
  • Element test karein: Service → Probe → Element Check
  • Cable near strain relief check karein
  • Test phantom ya water bath mein image quality check karein
  • حل / Fix

  • Correct frequency range set karein (C1-6: 1-6 MHz, abdomen ke liye 3.5 MHz typical)
  • Cable management improve karein — hanging weight se protect karein
  • Probe storage properly karein — hanging hook use karein
  • Engineer کب بلائیں

    Element failure ya probe face delamination = probe replacement needed. GE authorized repair center se estimate lein.

    Estimated Fix Time: 30 min diagnosis

    Difficulty: Intermediate

    🔧