🔬UltrasoundGE Logiq P7

GE Logiq P7 Touchscreen Issues

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Touchscreen respond nahi kar raha ya wrong position par respond kar raha hai.

ممکنہ وجوہات / Possible Causes

  • Touch calibration required
  • Screen surface contamination
  • Touch controller failure
  • Firmware bug
  • تشخیص / Diagnosis Steps

  • Screen surface thoroughly clean karein
  • Touch calibration test karein: Utility → System → Touch Panel → Calibrate
  • Firmware version check karein
  • Specific area fail ho rahi hai ya full screen
  • حل / Fix

  • Screen clean karein (microfiber cloth, no harsh chemicals)
  • Calibration re-run karein
  • Firmware update apply karein
  • Touch controller board replacement (engineer required)
  • Estimated Fix Time: 30–60 minutes

    Difficulty: Intermediate

    🔧