GE Logiq P7 Probe Not Recognized

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Logiq P7 probe connect karne ke baad "Probe Not Present" ya similar error aata hai.

ممکنہ وجوہات / Possible Causes

  • Probe connector dirty
  • Port lock mechanism not engaged
  • Probe software version mismatch
  • Port hardware fault
  • تشخیص / Diagnosis Steps

  • Probe properly lock hua hai check karein (click sound aana chahiye)
  • Dusra probe same port par test karein
  • Same probe dusre port par test karein
  • Software version aur probe compatibility check karein
  • حل / Fix

  • Connector clean karein
  • Lock mechanism properly engage karein
  • Software update se new probe models support mil sakta hai
  • Port hardware inspection (engineer required)
  • Engineer کب بلائیں

    Port-specific failure ya firmware compatibility issues ke liye GE service center se contact karein.

    Estimated Fix Time: 30–60 minutes