GE Logiq P5 Probe Error

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

GE Logiq P5 "Probe Fault" ya "Check Probe" error display karta hai.

ممکنہ وجوہات / Possible Causes

  • Probe connector pins bent ya corroded
  • Probe cable damaged (especially near connector)
  • Probe recognition chip faulty
  • Wrong probe selected in software
  • تشخیص / Diagnosis Steps

  • Probe connector visually inspect karein
  • Dusra compatible probe try karein
  • Software mein correct probe type selected hai check karein
  • Connector area moisture check karein
  • حل / Fix

  • Connector pins clean karein isopropyl alcohol se
  • Software mein probe manually select karein: Probe → Select → [Model]
  • Cable near connector kink check karein
  • Probe on different port try karein (P5 multiple ports hote hain)
  • Engineer کب بلائیں

    Probe connector assembly replacement ya internal port damage ke liye GE authorized engineer zaruri hai.

    Estimated Fix Time: 30–60 minutes