🔬UltrasoundGE Logiq P5

GE Logiq P5 Common Faults

By MedBazaar Technical Team · Troubleshooting Guide

GE Logiq P5 کے عام مسائل

1. Trackball Stuck/Not Moving

Trackball ball remove kar ke clean karein. Rollers par dust accumulate hota hai.

Fix: Ball nikalen, isopropyl se clean karein, rollers bhi clean karein.

2. Slow Performance

Windows XP slow ho jata hai time ke saath.

Fix: Disk cleanup, defragmentation (engineer se karwain safely).

3. Image Quality Degradation

Time gain compensation (TGC) check karein. Probe wear bhi cause ho sakta hai.

Fix: TGC reset, probe inspection.

4. Fan Noise (Loud)

Cooling fans dust se jam ho jate hain.

Fix: External air blast se vents clean karein. Internal cleaning engineer se.

5. Date/Time Reset on Every Boot

CMOS battery dead hai.

Fix: CMOS battery replacement (CR2032) — engineer required.

6. Monitor Flickering

Display cable loose ya inverter failing.

Fix: Engineer required.

Preventive Maintenance Tips

  • Monthly: External vents clean karein
  • Quarterly: Engineer se internal cleaning karwain
  • Annually: Full PM service schedule karein
  • Difficulty: Reference

    🔧