مسئلہ / Problem

Dental suction unit adequate suction provide nahi kar raha ya completely fail hai.

ممکنہ وجوہات / Possible Causes

  • Filter blocked
  • Canister full
  • Suction motor issue
  • Air leak in lines
  • Separator full
  • تشخیص / Diagnosis Steps

  • Canister/container level check karein — full to nahi
  • Filter condition check karein
  • Suction tip clear hai check karein
  • Any air leak sounds check karein (hissing)
  • Motor sound normal hai check karein
  • حل / Fix

    Canister Full:

  • Properly empty aur disinfect karein
  • Disposal protocol follow karein
  • Filter Blocked:

  • Filter clean ya replace karein
  • Frequency: daily rinsing, weekly replacement (disposable filters)
  • Air Leak:

  • All connections check karein
  • O-rings inspect karein — replace if worn
  • Tubing cracks check karein
  • Motor:

  • Motor overheating protection triggered: cool down 15 min
  • Motor replacement: engineer required
  • Infection Control

    Suction system proper decontamination daily zaruri hai. Approved enzymatic cleaner use karein end of day flush ke liye.

    Estimated Fix Time: 15–45 minutes

    Difficulty: Basic to Intermediate

    🔧

    Cant Fix It Yourself?

    Book a certified biomedical engineer in your city. Fast response, verified professionals.

    Find an Engineer →