Dental Chair Compressor Troubleshooting

By MedBazaar Technical Team · Troubleshooting Guide

مسئلہ / Problem

Dental compressor air pressure insufficient hai ya compressor noise kar raha hai.

ممکنہ وجوہات / Possible Causes

  • Air filter choked
  • Tank pressure low (leak)
  • Compressor motor issue
  • Safety valve triggered
  • Water in tank
  • تشخیص / Diagnosis Steps

  • Pressure gauge check karein — working pressure 5-6 bar hona chahiye
  • Air filter condition check karein
  • Any air leaks sound check karein (hissing)
  • Tank drain valve water check karein
  • حل / Fix

    Air Filter:

  • Filter element replace karein — quarterly ya as needed
  • Clean type filters: wash aur dry karein
  • Water in Tank:

  • Daily: drain valve khol ke water release karein
  • Water in lines: inline desiccant dryer install karein
  • Pressure Low:

  • Pressure switch setting check karein
  • Leaks fix karein (pipe thread tape use karein)
  • Compressor Noise:

  • Loose components tighten karein
  • Oil level check karein (oil-lubricated compressors)
  • Bearing noise: engineer required
  • Maintenance Schedule

  • Daily: drain tank water
  • Weekly: filter check
  • Monthly: pressure gauge verify
  • Annually: full service
  • Estimated Fix Time: Variable

    Difficulty: Intermediate

    🔧