مسئلہ / Problem

Scaler tip se water flow nahi ho rahi ya insufficient hai.

ممکنہ وجوہات / Possible Causes

  • Water supply disconnected ya low pressure
  • Filter/screen blocked
  • Internal tubing kinked
  • Water flow control valve issue
  • Tip orifice blocked
  • تشخیص / Diagnosis Steps

  • Water supply main valve open hai check karein
  • Water supply pressure adequate hai verify karein
  • Scaler water filter inspect karein
  • Tip orifice blocked to nahi
  • حل / Fix

    Water Supply:

  • Main water valve fully open karein
  • Water pressure check karein (adequate supply)
  • Filter:

  • Filter housing open karein
  • Screen/mesh filter rinse karein ya replace karein
  • Monthly cleaning schedule
  • Tip Orifice:

  • Thin wire se gently clear karein (tip orifice very small hota hai)
  • Dilute citric acid mein soak karein (calculus deposits remove)
  • Tubing:

  • Kinks straighten karein
  • Cracked tubing replace karein
  • No Water Safety

    Scaler ko water ke bina na chalain — tip aur tissue overheat ho sakte hain. Burn risk.

    Estimated Fix Time: 15–30 minutes

    Difficulty: Basic

    🔧